Micrometer Standard Bar Recall

Problem: Possible dimensional stability error in a small batch of 167 series 225mm Standard Bars.

Recently, it has been found that a possible dimensional error may occur in a small number of micrometer setting standard bars as supplied within the following micrometer sets or as supplied independently. As a result of this potential error, we hereby issue a recall notice and will replace the affected micrometer standard bars free of charge.

This recall affects only certain products produced within a defined time and within a range of serial numbers. The information below defines the affected units. Some of them are packaged within new micrometer’s sets while others have been sold individually.

As always, Mitutoyo strives to maintain the highest product quality and we thank you in advance for your understanding and cooperation in this matter. We apologise for any inconvenience this may cause to you, our valued customer.

 
Applicable Product
                         
   0223.jpg
・Product name      Micrometer standard bar (225mm)
・Model No.            MB-225
・Order No.            167-109
・Serial No.             Refer to the following table.
Serial No.  6047077 to 6047256
      6047370 to 6047371
6048270 to 6048332
6048349 to 6048465
6049007 to 6049148
6049150 to 6049187
*Applicable products are those delivered in and afterDecember, 2016.

Products equipped with the micrometer standard bar (225mm)

Product   Order No. Model No. 
Micrometer                103-146  OM-250
 103-146-10  OM-250
 103-914-40  OMST1-300B
 103-914-50  OMST1-300C
 103-915  OMST2-300
 103-915-10  OMST2-300
 104-136  OMC-300
 104-136A  OMC-300
 104-141  OMC-300
 104-141A  OMC-300
 111-124  SPM-250
 122-110  BLM-250
 123-110  GMA-250
 124-182  GMB-250
 143-110  OMP-250
 Digimatic Micrometer   293-255-30  MDC-250MX
 340-252-30  OMC-300MX
 Micrometer standard bar set  167-903  MBST-275


Fault description

Due to a fault in a heat treatment process the affected standard bars may incur a change in dimensional stability. 
As a consequence, if a micrometer zero point is adjusted using a faulty standard bar, the resulting adjustment will be adversely affected.

Remedy

We will replace the affected micrometer standard bar (225mm) free of charge. Please contact your Mitutoyo sales office

Inquiry

Please do not hesitate to contact your Mitutoyo sales office if you have any questions.